Part Number Hot Search : 
Z0803606 TM12864G 84BJC AS170 84BJC TC114Y 74HCT405 MRF24J40
Product Description
Full Text Search
 

To Download AQ6370 Datasheet File

  If you can't view the Datasheet, Please click here to try to view without PDF Reader .  
 
 


  Datasheet File OCR Text:
 www..com
AQ6370
OPTICAL SPECTRUM ANALYZER
World Class Optical Performance & Flexibility*
* * * * High wavelength resolution: 0.02 nm Wide close-in dynamic range Multimode fiber test capability (up to GI 62.5/125m) Pulsed light measurement capability
* In the diffraction-grating-based optical spectrum analyzer industry as of January 2006
Improves Measurement Throughput
* Fast measurement * Fast data transfer and storage
Expedites Development of Automated Test Systems
* Supports GP-IB, RS-232C, and Ethernet interfaces * Compatible with SCPI and supports AQ6317 series remote commands * Built-in simple macro programming function
Enhances User Friendliness
* Supports mouse and keyboard operation (USB & PS/2) * Trace zoom capability
Facilitates Frequent Data Handling
* Large internal user memory (5000+ traces) * Supports USB 1.1 compatible large external storage devices
Includes Wavelength Calibration Source
Bulletin 7353-01E
www..com
100x 10x
100x 10x
31
SWEEP
Improves Measurement Throughput 10x
FAST MEASUREMENTTRACES
New High Sensitivity Mode
5000+
FAST DATA TRANSFER AND STORAGE
5x 100x 10x
Through newly developed noise reduction techniques, the AQ6370 can complete a measurement faster even with a high sensitivity setting. This is imperative when measuring weak signals. A new sensitivity mode makes it up to five times faster*.
100x 100x
100x 100x
100x 10x 10x 10x
ETHERNET provides up to one hundred times 31
faster data transfer speed than the GP-IB data transSWEEP SWEEP fer*.
31
USB
100x 10x
31
Seamless Sweep
SWEEP
100x 10x 10x 5x 5x
GP-IB1 3 provides up to ten times faster data transfer
speed than the conventional GP-IB*. SWEEP
TRACES TRACES
5000+ 5000+
With an improved gain control system in the electrical amplifier circuit, sweep speed has been increased for the signals which require multiple amplifier gain setTRACES tings to complete an entire trace. The new system achieves up to ten times faster sweep*.
USB provides up to ten times faster data recording/
retrieving speed to a removable memory device than the floppy disk*.
5000+
TRACES USB
5000+
SATISFACTION WITH SINGLE MEASUREMENT
Quick Key Response 5x USB 5x Applying a faster microprocessor and new algorithms, 100x time100x response 31 100x key100x time and initial hardware setup before a sweep is drastically reduced. The new system
achieves up to one hundred times faster response*.
31
SWEEP
SWEEP
With the increased number of data sampling points (max 50,001), the AQ6370 can measure a wider wavelength range (span) with higher resolution. Single sweep can complete the test that took three sweeps in the past*.
*In comparison with AQ6317 series optical spectrum analyzer.
USB
10x
10x
Expedites Development of Automated Test Systems
The AQ6370 is equipped with GP-IB, RS-232C, and Ethernet (10/100BASE) interfaces to be connected with an external PC for remote access and building an automated test system. It is compatible with a standardized programming language and supports AQ6317 commands for easy programming. Macro Program is a useful built-in function for making a simple auto test program.
TRACES
5000+
10x
10x
TRACES
5000+
5x
USB
5x
USB
\
Image of OSA upgrading on the existing test system Host (Controller)
COMPATIBLE WITH SCPI
The SCPI is an ASCII text based standard code and format that conforms to IEEE-488.2. The standard remote commands of the AQ6370 are compatible with SCPI. It can be used with most computer test languages and test application software.
AQ6317 EMULATION MODE
The AQ6317 series is Yokogawa's best-selling optical spectrum analyzer. Since there are so many existing users all over the world, the AQ6370 supports private remote programming codes and formats of the AQ6317 series to make it easier for users to upgrade their current test environment.
Note. some commands may not be compatible due to changes in specifications and functions.
Measurement Conditions Control of external equipment Execution of a measurement Control of external equipment GPIB Port 2, Ethernet or RS-232C Peripheral instruments (slave)
MACRO PROGRAMMING
Macro programming enables user to easily create test procedures by recording the user's actual key strokes and parameter selections. One program can contain up to two hundred program lines, and programs can be called as subroutines to make the main program simple. Moreover, macro programs can control external equipment through Ethernet, RS-232C and GP-IB port (GP-IB2). Therefore, an external PC is not required to build a simple auto test system.
Calculation between traces
Example of a simple auto test system with Macro program
LabVIEW (R) DRIVER SUPPORT
LabVIEW(R) is a popular test application software. Yokogawa can provide users with LabVIEW drivers for the AQ6370.
2
www..com
Enhances User Friendliness
INHERITS PROVEN OPERABILITY
The AQ6370 inherits front panel operation and function key assignments from our conventional models that are proven intuitive and easy to use by our multitude of users.
Parameter entry window appears on the screen with the e n t r y k ey s fo r mouse operation when a parameter is selected.
MOUSE & KEYBOARD OPERATION
You can easily operate the AQ6370 with only a mouse instead of the front panel keys. Measuring conditions displayed on the trace screen can be changed directly by the mouse pointer. The keyboard is useful for entering labels and file names.
Direct parameter entry Arrow pointer transforms into finger pointer when you move it to parameters. Front panel keys
[
The functions of panel keys appear by clicking right mouse button.
TRACE ZOOM CAPABILITY
Trace zoom makes it possible to change display conditions, such as center wavelength and span, by clicking and dragging the mouse to select the designated area to enlarge. No need to make another sweep to refresh the display conditions anymore. The analysis function can be performed in the enlarged area.
1 5
s
s
click
Original waveform in measured wavelength range
1 click
s
\
5 click
s
Enlarged waveform
In the Analysis Screen, traces can also be zoomed.
When a line item on the analysis result table is selected, the display area is automatically shifted and the corresponding signal is centered in the display area. This makes it easier to verify the accuracy of the analyzed data.
drag
Overview Window
Once the trace is zoomed in, the overview window appears and shows the entire trace. By dragging the zoomed area border in this window with the mouse, the display area in the main trace window can be modified.
Overview 100x window 100x
31
SWEEP Operation in Overview window
00x 0x
10x Facilitates Frequent Data Handling + 5000 31 10x
LARGE INTERNAL MEMORY
TRACES
SWEEP
SUPPORTS USB 1.1 MEMORY
TRACES
5000+
The AQ6370 has a 128 MB user area in the internal memory that can save test setups, waveforms, analysis results, and macro program files. It is large enough to save more than five thousand traces.
5x
USB
Thumbnail file preview
makes it easy to find a particular file out of thousands of files. (The Thumbnail file preview function also works with an external memory.)
Thumbnail file preview
USB
The AQ6370 has two USB 1.1 compatible interfaces. They support large size removable memory devices such as Flash ROM and hard disk drives (HDD). Removable static memory of more than 1 GB and USB compatible external HDDs of more than 40 GB in capacity are commercially available.
(as of January 2006)
3
www..com
World Class Optical Performance and Flexibility*
EXCELLENT SIGNAL SEPARATION
The AQ6370 uses a double-pass monochromator structure to achieve high wavelength resolution (0.02 nm) and wide close-in dynamic range (70 dB). Thus, closely allocated signals and noise can be separately measured. OSNR measurement of 50 GHz spacing DWDM transmission systems and EDFA evaluation with multiple wavelength sources can successfully be performed.
AQ6370
4
WIDE MEASUREMENT RANGE
Power: +20 dBm to -90 dBm
The AQ6370 can measure optical power as high as +20 dBm, which enables direct measurement of high power sources such as optical amplifiers and pump lasers for Raman amplifiers. Measurement sensitivity can be chosen from seven categories according to test applications and measurement speed requirements.
Wavelength: 600 nm to 1700 nm
The AQ6370 covers not only telecommunication wavelengths, but also the visible light wavelength region which is used for home electronics, medical, and industrial material applications. SENS
70dB +20dBm
Level sensitivity settings; NORM Hold, NORM Auto, NORM, MID, and HIGH (1, 2, and 3), that correspond to sensitivity from 60dBm through -90dBm.
-90dBm Peak0.4nm
Sample waveforms of high power (+20dBm) and Low power (-90dBm) signal measurements HIGH
Close-in Dynamic Range 70dB at peak 0.4nm, resolution setting 0.02nm (typical)
High dynamic range mode settings; CHOP and SWITCH are modes to obtain a better dynamic range by reducing an influence of stray-light. \ They are applicable in MID and HIGH mode.
MULTIMODE FIBER TEST CAPABILITY
The AQ6370 uses a free space input structure that can handle up to GI 62.5/125 multimode fiber. Multimode fiber is commonly used in high speed Ethernet network, such as GE-PON. The free space input is also beneficial for measurement repeatability as insertion loss variation at the input connector is smaller than the other input type which has an optical fiber inside the monochromator.
PULSED LIGHT MEASUREMENT CAPABILITY
The AQ6370 can catch the peak power of a pulsed signal using PEAK HOLD or using an externally provided trigger to synchronize with the measured signal. It can be applied to the transmission loop testing of telecommunication systems, and also to the low power measurement at the early stage of laser chip development since it works in the high sensitivity modes.
CW light measurement mode Pulsed light measurement mode
Sample waveform of an 850nm laser with a multimode fiber (62.5/125mm)
Structure of the free space monochromator input
Sample waveform of a pulsed light signal
* In the diffraction-grating-based optical spectrum analyzer industry as of January 2006
Real-Time Remote Operation
Optional Application Software (Under development )
AQ6370Viewer is PC application software designed to work with Yokogawa's AQ6370 Optical Spectrum Analyzer. It has exactly the same user interface and functions as the AQ6370 so that you can easily display and analyze waveform data acquired by the AQ6370.
Menu bar
for Monitoring and Troubleshooting
Remote control function
The remote control allows you to control AQ6370 Optical Spectrum Analyzer from anywhere on the Ethernet network. This gives you the sensation of setting measurement conditions and analysis parameters, and executing measurement on an actual unit. Because of fast data transfer speed of Ethernet, measurement data can be updated in real time. It is useful for diagnosis of trouble in production lines, periodical observation of long term tests in the lab, and other applications.
Note. the data update speed varies depending on network performance and conditions.
ETHERNET
Status bar
Panel key menu Image of remote access via Ethernet using the remote control function of AQ6370 Viewer
Screen example of AQ6370 Viewer
www..com
Advantageous Functions
PASS/FAIL AUTO TEST FUNCTION
The waveform can automatically be judged to PASS or FAIL against specified conditions. Using Template function upper and/or lower limits for the assessment can be set and measured waveform can automatically be compared. It is an effective way to reduce time and human error in assessment, especially for production line tests. The template data can be created and stored in the AQ6370. It can be edited using a spread sheet on an external computer as well.
AQ6370
7 TRACES & CALCULATION FUNCTIONS
The AQ6370 has seven individual traces for measurement data. Some traces can be used for calculations (two-trace subtraction and addition), MAX/MIN hold, averaging, and curve fitting.
Trace G can calculate between calculated traces (trace C and F)
TRACE G TRACE F TRACE E TRACE D TRACE C TRACE B TRACE A
Trace F can calculate between trace D and E Trace C can calculate between trace A and B
Example of trace calculations
Trace can be selected directly by the right-click in the trace window.
Enhanced Curve Fit function
Sample waveform of PASS/FAIL test Screen example of Template editor
13 AUTO ANALYSIS FUNCTIONS
Thirteen types of built-in analysis functions for popular applications can be selected by using Analysis function key on front panel or by mouse operation. The functions automatically perform designated analyses and provide results. The results can be saved in a storage device. * * * *
A B
The Curve fit function is an approximation technique to exclude an influence of noise and/or signal overlapping on the waveform. Curve fit method can be chosen from Gaussian, Lorenz, 3rd Poly, 4th Poly, and 5th Poly. Curve fit can be applied to an individual trace. Fitting area can be set by line markers (L1 and L2).
xAnalysis mode \ When WDM or WDM-NF analysis is executed, one of the curve fit methods is used for a baseline measurement. The curve fit method can be specified in the parameter window of each analysis.
A B
WDM analysis (OSNR) WDM-NF analysis (EDFA) DFB-LD analysis Filter analysis, etc. Analysis Parameter Window
All parameters for an individual analysis are displayed in a dialog box for easy setup.
[ xExample of curve fitting in selected areas
Noise level of amplified WDM signal can be estimated by using the curve fit in selected areas, even if noise between signals cannot be seen due to a limit of signal separation performance.
Fitting area
xMarker fitting \ The fitting curve can also be applied to the marker positions set by user. For example, this can be used to measure a modulated signal spectrum.
Connectivity
FRONT PANEL
OPTICAL CONNECTORS
(USER EXCHANGEABLE )
REAR PANEL
USB 1.1 INTERFACE
SERIAL
(RS-232C)
TRIGGER IN (TLS Sync Sweep) TRIGGER OUT (TLS Sync Sweep) ANALOG OUT
The AQ6370 adopts a universal type optical connector system for optical input and calibration output enabling direct coupling to major optical connector types. The connectors can be replaced by users.
Mouse
KEYBOARD (PS/2) VIDEO OUT(SVGA)
Keyboard
Memory
GP-IB (IEEE488.1/488.2)
GP-IB 1 : Allows remote control from your PC GP-IB 2 : AQ6370 controls external instruments
HDD
ETHERNET (10/100BASE-TX)
5
www..com
Applications
SIMPLE EDFA TEST
The ASE interpolation method is used to measure gain, NF, and key parameters for optical fiber amplifier evaluation. With WDMNF analysis function, up to 1024 channels of multiplexed signals can simultaneously be tested. An ASE level for NF measurements is calculated by using a curve-fit function for each WDM channel. The curve-fit and source spontaneous emission (SSE) suppress function enhance a measurement accuracy.
AQ2200-111 DFB-LD modules Optical amplifier Multiplexer Optical attenuator/ switch module AQ6370 AQ6370 WDM coupler
AQ6370
6
AUTOMATED EDFA TEST
In conjunction with DFB-LDs, optical attenuator and optical switches, an EDFA auto test system can easily be established. This system does not require manual reconfiguration of optical paths and adjustments of EDFA input power, and that improves measurement throughput and avoids a human error. AQ2200 series is a modular system suitable for building such a system.
AQ2200-111 DFB-LD modules
DUT WDM waveform before/after amplification by EDFA Example of automated EDFA test configuration
TRANSCEIVER / LD TEST
In conjunction with bit error rate test (BERT) equipment, the AQ6370 can measure the center wavelength and spectral width of transceivers and LD modules. Various built-in analysis functions, such as DFB-LD, FP-LD (VCSEL), and LED facilitate test process.
BERT(PPG) Transceiver AQ6370
WDM OSNR TEST
AQ6370's wide close-in dynamic range allows accurate OSNR measurement of DWDM transmission systems (up to 50 GHz spacing). The built-in WDM analysis function analyzes the measured waveform and shows peak wavelength, peak level and OSNR of WDM signals up to 1024 channels simultaneously. The curve fit function is used to accurately measure noise levels.
AQ2200-601
Example of analyzed data table
Tx1 Tx2 Tx3 Tx4
1 2 3 4 n
Optical amplifier
MUX
AQ6370
Example waveform and analysis result of transceiver output
TxN
Example of WDM OSNR analysis
PASSIVE COMPONENT TEST
Wide dynamic range measurement using a synchronous wavelength sweep function* of a tunable laser source and optical spectrum analyzer is suitable for evaluating passive devices and components with a high crosstalk ratio. The tunable laser source emits a single wavelength, and the AQ6370's filter characteristics cuts source spontaneous emission and scattered light. Thus, this system can achieve wide dynamic range over 70 dB.
* TLS SYNC function. It supports AQ4320, AQ4321 and AQ2200-136 Tunable Laser Source.
In conjunction with a white light source, an ASE light source or other broadband light source, you can simply perform evaluation of passive devices such as WDM filters and FBG. The AQ6370's superb optical characteristics enable higher-resolution and wider dynamic range measurements. The built-in optical filter analysis function simultaneously reports peak/bottom wavelength, level, crosstalk, and ripple width.
AQ2200-136 TLS module
Synchronous signal AQ6370 GP-IB AQ4305 AQ6370
DUT(FBG, WDM filter, etc. ) DUT(Optical multiplexer) Example of AWG measurement Example of CWDM filter measurement
www..com
Specifications
Applicable fiber Measurement wavelength range 1) Span 1) Wavelength accuracy 1), 2), 3) SM (9.5/125 m), GI (50/125 m, 62.5/125 m) 600 to 1700 nm 0.5 nm to full range and zero span 0.02 nm (1520 to 1580 nm) 0.04 nm (1450 to 1520 nm, 1580 to 1620 nm) 0.1 nm (Full range) 0.01 nm (1520 to 1580 nm) 0.02 nm (1450 to 1520 nm, 1580 to 1620 nm) 0.005 nm (1 min.) 101 to 50001 0.02, 0.05, 0.1, 0.2, 0.5, 1.0 and 2.0 nm 5 % (1450 to 1620 nm, resolution setting: 0.1 to 2.0 nm,resolution correction: ON, measurement data point setting: AUTO) NORM_HOLD, NORM_AUTO, NORMAL, MID, HIGH1, HIGH2 and HIGH3 Switch (Sensitivity: MID, HIGH1, HIGH2 and HIGH3) CHOP (Sensitivity: HIGH1, HIGH2 and HIGH3) -90 dBm (1300 to 1620 nm, resolution: 0.05 nm or wider, sensitivity: HIGH3) -80 dBm (1000 to 1300 nm, resolution: 0.05 nm or wider, sensitivity: HIGH3) -60 dBm (600 to 1000 nm, resolution: 0.05 nm or wider, sensitivity: HIGH3) 0.4 dB (1310/1550 nm, input level: -20 dBm, sensitivity: MID, HIGH1, HIGH2 and HIGH3) 0.05 dB (Input level: -50 to +10 dBm, sensitivity: HIGH1, HIGH2 and HIGH3) 0.1 dB (1520 to 1580 nm) 0.2 dB (1450 to 1520 nm, 1580 to 1620 nm) +20 dBm (Per channel, full span) +25 dBm (Total safe power) 45 dB (0.2 nm from peak at 1523 nm, resolution: 0.05 nm) 62 dB (0.4 nm from peak at 1523 nm, resolution: 0.05 nm) 40 dB (0.2 nm from peak at 1523 nm, resolution: 0.1 nm) 57 dB (0.4 nm from peak at 1523 nm, resolution: 0.1 nm) 0.05 dB (1550/1600 nm) 0.08 dB (1310 nm) Per sensitivity mode; NORM_AUTO: 0.5sec NORMAL: 1 sec MID: 2 sec HIGH1: 5 sec HIGH2: 20 sec HIGH3: 75 sec Function Automatic measurement Setting of measuring conditions Macro program function (64 programs, 200 steps) Center wavelength setting, Span setting, Measurement data point setting, Wavelength resolution setting, Sensitivity setting, High dynamic range mode setting, Averaging number setting (1 to 999 times), Automatic measuring conditions setting, Sweep between line markers, zero span sweep, Automatic measurement data point setting, Pulse light measurement, External trigger measurement, Sweep trigger, Sweep status output, Analog output, TLS s y n c h r o n i ze d swe e p, A i r / va c u u m wave l e n g t h measurement, Pass/Fail judgment with template Level scale setting (0.1 to 10 dB/div. and linear), Vertical sub scale setting (0.1 to 10 dB/div. and linear), Reference level and position setting, Vertical division number setting (8, 10 or 12), Frequency horizontal scale display, Horizontal scale zoom in/out display, Measurement condition display, Noise mask, Data table, Label, Split display, % display, dB/nm (power spectral density) display, dB/km display, Template display, 7 independent traces, Write/Fix setting, Display/Blank setting, Max./Min. hold, calculation between traces, Roll (Sweep) averaging (2 to 100 times), Normalized, Curve fit/Peak curve fit/Marker curve fit, Trace copy function, Trace clear function Delta marker (Max. 1024), Vertical/Horizontal line marker, Peak search, Next peak search, Bottom search, Next bottom search, Auto search, Search between horizontal line markers, Search in the zooming area Spectral width (threshold, envelope, RMS, Peak RMS, notch), WDM (OSNR) analysis, EDFA-NF analysis, Filter peak/bottom analysis, WDM filter peak/bottom analysis, DFB-LD analysis, FP-LD analysis, LED analysis, SMSR analysis, Power analysis, PMD analysis, Pass/Fail judgment with template, Auto analysis, Analysis between horizontal line markers, Analysis in the zooming area Self optical alignment function with built-in light source, Self wavelength calibration function 64 Traces, 64 programs, 3 template lines Max. 128MByte USB storage (memory/HDD) Capability, FAT32 format CSV(text)/Binary, BMP/TIFF GPIB, RS-232C and Ethernet (TCP/IP) AQ6317 series compliant commands (IEEE488.1) and IEEE488.2 full support GPIB x2 (standard/controller), RS-232C, Ethernet, USB1.1 x2, PS/2 (keyboard), SVGA output, Analog output port, Trigger input port, Trigger output port Free space optical input: Requires AQ9447 (*) connector adapter PC contact built-in light source output: Requires AQ9441 (*) Universal adapter Built-in high-speed thermal printer (Factory option) 10.4-inch color LCD (Resolution: 800 x 600) 100 to 240 VAC, 50/60Hz, approx. 150VA Operating temperature: +5 to +35 C Storage temperature: -10 to +50 C Humidity: 80 %RH or less (no condensation) Approx. 426 (W) x 221 (H) x 459 (D) mm, Approx. 27kg (without printer option)
AQ6370
Wavelength linearity 1), 2), 3) Wavelength repeatability 1), 2) Measurement data point Wavelength resolution setting 1), 2) Resolution accuracy 1), 2), 3)
Display
Level sensitivity setting High dynamic range mode
Traces
Level sensitivity 2), 4), 5), 7)
Marker/Search
Level accuracy 2), 4), 5), 6) Level linearity 2), 4) Level flatness 2), 4), 6) Maximum input power 2), 4) Safe max. input power 2), 4) Close-in dynamic range 1,) 2), 9)
Analysis
Other Data storage Internal memory Internal storage External File type Remote control
Interface
Polarization dependency 2), 4), 6) Sweep time 1), 7), 8)
Category
Optical connector
Printer Display 11) Power requirement Environmental conditions
Dimensions
Unit : mm (approx. inch)
Dimensions and mass 10)
REAR VIEW
14.8 (0.58) 13 (0.51) 426 (16.77) 13 (0.51)
(1.26) 32
427 (18.22)
(1.26) 32
221 (8.70)
Note: 1) Horizontal scale: wavelength display mode 2) At 235 C, with 10/125 m single mode fiber, after 2 hours of warm-up, after optical alignment with built-in reference light source 3) After wavelength calibration with built-in reference light source 4) Vertical scale: absolute power display mode, resolution setting: 0.05 nm or wider, resolution correction: OFF 5) With 10/125 m single mode fiber (B1.1 type defined on IEC60793-2, PC polished, mode field diameter: 9.5 m, NA: 0.104 to 0.107) 6) Temperature condition changes to 233 C 7) High dynamic range mode: OFF, pulse light measurement mode: OFF, TLS sync sweep: OFF, resolution correction: OFF 8) Span: any 100 nm or less, measurement data point: 1001, average number: 1 9) High dynamic range mode: CHOP or SWITCH, resolution correction: OFF 10) Excluding feet and handles 11) Liquid crystal display may include few defective pixels (within 0.002% with respect to the total number of pixels including RGB). There may be few pixels on the liquid crystal display that do not emit all the time or remains ON all the time. Note that these are not malfunctions.
Standard Accessories
Name Power cable User's manual (1set) Q'ty 1 1
20 (0.79)
7
www..com
Factory Installed Options
BUILT-IN PRINTER
An optional built-in thermal printer is provided to instantly print out a screenshot of the AQ6370's display, analysis results, a marker list and a macro program list.
Accessory: printer roll paper (1 roll)
OPTICAL CONNECTOR ADAPTERS
For optical input port
AQ9447 Connector Adapter /FC, /SC, /ST
For calibration output port
AQ9441 Universal Adapter /RFC, /RSC, /RST
Ordering Information
Model and Suffix Codes
Model 735301 Power cable Suffix Codes -D -F -G -Q -H Factory Installed Options -M /FC /SC /ST /RFC /RSC /RST /B5 Descriptions Optical Spectrum Analyzer AQ6370 Power cord (UL3P) Power cord (CEE-C7) Power cord (SAA-3P) Power cord (BS3P Rectangular) Power cord (BS3P Round) Power cord (UL3P with 3P/2P converter) AQ9447(FC) Connector adapter for optical input AQ9447(SC) Connector adapter for optical input AQ9447(ST) Connector adapter for optical input AQ9441(FC) Universal adapter for calibration output AQ9441(SC) Universal adapter for calibration output AQ9441(ST) Universal adapter for calibration output Built-in thermal printer Printer roll paper B9988AE AQ9441 Universal adapter Connector type 813917321 -FCC -SCC -STC
Accessories (Optional)
Name Model AQ9447 Connector adapter 810804602 Connector type Suffix codes -FCC -SCC -STC Specifications For optical Input port FC type SC type ST type For calibration output port FC type SC type ST type 10 m roll, 10 rolls/1 unit
Related Products
Optical Spectrum Analyzer White Light Source Tunable Laser/DFB-LD Sources
AQ6319
AQ4305
AQ2200 Series
Superior Performance
Broadband
Modular Platform
Microsoft, MS, and Windows are registered trademarks or trademarks of Microsoft Corporation in the US and other countries. LabVIEW is a U.S. registered trademark of National Instruments. Other company names and product names appearing in this document are the registered trademarks of their respective comparnies. "Typical" or "typ." in this document means "Typical value", which is for reference, not guaranteed specification.
Note
q Before operating the product, read the user's manual thoroughly for proper and safe operation. q If this product is for use with a system requiring safeguards that directly involve personnel safety, please contact the Yokogawa sales offices.
YOKOGAWA ELECTRIC CORPORATION Communication & Measurement Business Headquarters /Phone: (81)-422-52-6768, Fax: (81)-422-52-6624 E-mail: tm@cs.jp.yokogawa.com YOKOGAWA CORPORATION OF AMERICA Phone: (1)-770-253-7000, Fax: (1)-770-251-6427 YOKOGAWA EUROPE B.V. Phone: (31)-33-4641858, Fax: (31)-33-4641859 YOKOGAWA ENGINEERING ASIA PTE. LTD. Phone: (65)-62419933, Fax: (65)-62412606
Subject to change without notice.
[Ed : 01/b] Copyright (c)2006 Printed in Japan, 602(KP)
MS-16E


▲Up To Search▲   

 
Price & Availability of AQ6370

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X